QUALITY CONTROL TESTS FOR SUTURES AND LIGATURES


1)    GAUGE TEST:
For checking the diameter of the catgut, the dial reading type gauge is used. The diameter is taken in a relative humidity between 60%-80% and temperature range 16-21oC.
The finest size is 1, which is used for ophthalmic work and the largest size is 10, which is used for other surgical procedures.

2)    LENGTH:
This is measured immediately after the removal of suture from the container without stretching. The length should not be less than 90% of the stated length on the label.

3)    TENSILE STRENGTH:
It is done by means of machine in which (Load necessary to rupture the GUT) is measured. The test is performed on straight and knotted sample. The suture is tested within 15 minutes of the removal from the container at a temperature 10-21oC and relative humidity 60%-80%.

4)    STERILITY TEST:
A medium is prepared which can detect the presence of aerobic and anaerobic organisms. If after an incubation period of 14 days, no growth of the micro-organisms is found in any tube, the sample passes the test.
If the growth occurs, another sample is taken and the test is repeated. If the growth occurs, the catgut fails in sterility test.